More concerning - can anyone explain why there is such a variation in the results from the DC powered unit vs. the TENG-powered one? The graph at the bottom of the report shows a difference of 30-50ppm between both units when they are sitting side by side on the bench.
I skimmed the original article and it only mentions the graph and says that it's "comparable to DC powered unit". I'm guessing < 100ppm difference is somewhat acceptable?
That's in the normal range of accuracy of modern CO2 sensors, for example SCD40 from Sensirion is described with an accuracy of ±50.0 ppm ±5.0 %m.v.