I don't think that's true at all. Capnography, the measure of carbon dioxide partial pressure is wholly separate from pulseox:

> Pulse oximeters have some limitations. They can only employ light at two wavelengths. Thus the devices can only distinguish between hemoglobin and oxygenated hemoglobin. When carboxyhemoglobin and methemoglobin are also present, there are two additional wavelengths required for differentiation. In the presence of elevated carboxyhemoglobin levels, pulse oximetry overestimates the true saturation of oxygen as carboxyhemoglobin binds with a higher affinity than oxygen. In the case of carbon monoxide poisoning, the absorbance spectrum of carbon monoxide is very similar to hemoglobin, which results in a falsely high level of oxygen (overestimation of oxygen saturation) ...

https://www.ncbi.nlm.nih.gov/books/NBK539754/


Replies

benj111today at 9:50 AM

>Pulse oximeters have some limitations. They can only employ light at two wavelengths

Why only 2?

show 1 reply